Ara

    Nakanishi Inc.NSK Logo

    Su Tedarik Devresi

    Su Tedarik Devresi

    If you experience any one of the following failures, try step-by-step troubleshooting described below.

    • Insufficient water supply
    • Irregular water supply
    • No mist spray
    • No water at all

    Identifying the Blockage

    • Hose
      Disconnect the turbine from the hose to see whether there is sufficient water supply from the hose. If there is not enough water supply, the most likely causes are clogging of the hose, poor connection, lower water pressure or a malfunction of the dental unit.
    • Coupling Joint
      Connect a turbine body (without a cartridge) to the coupling and run a test. If there is not sufficient water supply, you'll know the blockage is within the turbine body.
      • a. Clogging caused by water scale
      • b. Insufficient water pressure caused by a loose coupling
    • Coupling Joint
      Connect a turbine body (without a cartridge) to the coupling and run a test. If there is not sufficient water supply, you'll know the blockage is within the turbine body.
      • a. Clogging caused by cruds around the water hole
      • b. Clogging within the water circuit of turbine body
    • Cartridge
      If you see sufficient water supply at Step 3 above, then clogging should be in the cartridge. a. Clogged water spray nozzle